Magnetic cephalometric diagnostics for orthodontics DGKFO

noXrayCeph Magnetic cephalometric diagnostics informations

INTRODUCTION: There is still a deviation between in vivo and in vitro examination in cephalometric examination. To diminish these differences, patients were measured with a fixed head device, noXrayCeph, which is moveable. The fixation to the moving patient should not make any difference in an unmoving in vitro examination. Advantages of the direct scanning with noXrayCeph®:

Diagnosis of the right and left side (asymmetry), secure scanning of anatomical points with the ultra-light sensor stylus, scanning of relevant points with low pressure charge, freedom of movement of the head, which enables an orthodontic diagnosis of moving patients, rapid scan and immediate display of the result in graphics and values, multiple scans make it possible to check the treatment progress.

MATERIALS AND METHOD: To evaluate the accuracy of the noXrayCeph-Device, a patient and an anatomic skull were measured 10 times by different trained and untrained observers. The device from the noXrayCeph AG, CH-Meilen is a magnetic scanner with a carbon headcap, including three magnetic coils. The stylus is measures the x, y and z coordinates with a technical error of 0.22 mm.

RESULTS: The three-dimensional data produces a picture of the skull, reduced to the 76 points of orthodontic importance. The accuracy of the measurements from four different observers were calculated using the Wilcoxon-test. Intraobserver test was significant at p < 0.01 for the ANB, the position of MX 11 and 21 point to A-point and the incisors angulation. The interobserver test was significant at the 0.05 level.

CONCLUSIONS: The noXrayCeph device is a reproducible with a low intra- and inter- observer error. Reproducibility can be measured on one patient, because cephalometric diagnostics is possible without radiographs. Due to fast measurements the device seems to be useful in clinics, where different observers examine the patients. The low weight of the carbon head cap allows movements of the patient’s head, because the coordinates move 1:1 with the patient.